Neonatal mechanical ventilation is a cornerstone in the management of respiratory failure in newborns (NBs), particularly in preterm infants, in whom pulmonary and neurologic increases their vulnerability to ventilation-induced lung injury. At present, the primary objective of respiratory support is to ensure adequate ventilation while minimizing lung and brain injury through lung-protective strategies and careful oxygen titration.
Noninvasive ventilation (NIV) is currently the first-line approach for both initial respiratory support and post-extubation care, as it reduces the need for invasive ventilation and the incidence of ventilator-induced lung injury. Multiple modalities and devices are available, and their selection should be individualized according to the underlying disease, gestational age, and clinical response. The selection of the appropriate interface and ventilatory settings and patient-ventilator synchronization are of utmost importance. Noninvasive ventilation failure should be preceded by systematic optimization and requires well-defined criteria for timely therapeutic escalation.
Conventional mechanical ventilation continues to be necessary in more severe cases. Its use should be synchronized and volume-targeted whenever possible, as this strategy has been shown to reduce major complications such as pneumothorax, intraventricular hemorrhage, and the composite outcome of death or bronchopulmonary dysplasia. High-frequency oscillatory ventilation serves as a rescue strategy in refractory cases and selected clinical condition, relying on the maintenance of lung recruitment with extremely low tidal volumes.
Finally, ventilatory management of the most common neonatal respiratory conditions—such as respiratory distress syndrome, meconium aspiration syndrome, air leak syndrome, and bronchopulmonary dysplasia—requires a dynamic, individualized approach focused on lung protection from the earliest hours of life.
La ventilación mecánica neonatal constituye un pilar fundamental en el manejo de la insuficiencia respiratoria del recién nacido (RN), especialmente del RN prematuro, cuya inmadurez pulmonar y neurológica incrementa su vulnerabilidad a la lesión inducida por ventilación. El objetivo principal del soporte respiratorio actual es proporcionar una ventilación eficaz minimizando el daño pulmonar y cerebral, mediante estrategias de ventilación protectora y un uso juicioso del oxígeno.
La ventilación no invasiva (VNI) es, en la actualidad, la primera línea tanto en el soporte respiratorio inicial como tras la extubación, reduciendo la necesidad de ventilación invasiva y la incidencia de lesión pulmonar inducida por la ventilación. Existen múltiples modalidades y dispositivos, cuya selección debe individualizarse según la patología, la edad gestacional y la respuesta clínica, siendo clave una correcta elección de interfase, parámetros y sincronización. El fracaso de la VNI requiere una optimización sistemática previa y criterios bien definidos para la escalada terapéutica.
La ventilación mecánica convencional sigue siendo necesaria en situaciones de mayor gravedad. Su aplicación debe ser siempre sincronizada y con volumen garantizado, lo que ha demostrado reducir complicaciones graves como neumotórax, hemorragia intraventricular y la combinación de muerte o displasia broncopulmonar. La ventilación de alta frecuencia oscilatoria representa una herramienta de rescate en casos refractarios y en determinadas patologías específicas, basada en el mantenimiento del reclutamiento pulmonar y volúmenes muy bajos.
Finalmente, el manejo ventilatorio de los cuadros clínicos más frecuentes —como el síndrome de distrés respiratorio, la aspiración meconial, la fuga aérea y la displasia broncopulmonar— exige una aproximación dinámica, individualizada y orientada a la protección pulmonar desde las primeras horas de vida.
Neonatal mechanical ventilation (MV) is an essential tool in the management of neonates with respiratory failure, especially in preterm (PT) infants and those with severe pulmonary or systemic disease.1 Its correct implementation requires a thorough understanding of neonatal respiratory physiology, lung mechanics, and, in particular, the specific pathophysiology of each clinical condition, in order to select the most appropriate ventilatory strategy and provide effective and safe respiratory support.2
Newborn infants, especially those born preterm, are particularly vulnerable to the effects of mechanical ventilation. Their respiratory system, including its vasculature, is still structurally and functionally underdeveloped, while their central nervous system is undergoing a critical phase of organization and differentiation. Both of these factors greatly increase their vulnerability to various external stressors.1–3 In this context, both mechanical ventilation and oxygen toxicity can contribute to the development of ventilation-induced lung injury (VILI).1,2,4 Furthermore, changes in cerebral blood flow and fluctuations in carbon dioxide levels and acid-base balance can contribute to neurologic complications, such as intraventricular hemorrhage or periventricular leukomalacia, which may have significant long-term consequences.1–3 Therefore, the approach to respiratory support should focus on protective mechanical ventilation strategies, with careful adjustment of ventilatory settings and oxygen titration. In this regard, the current evidence recommends maintaining oxygen saturations between 90% to 95% in infants born before 28 weeks of gestation until 36 weeks of corrected gestational age.1,2,4
Understanding these particular pathophysiological characteristics is essential for optimizing the use of mechanical ventilation and minimizing the risk of lung and brain injury in newborns.1–3
Noninvasive ventilationIntroductionNoninvasive ventilation (NIV) delivers constant pressure to the airways, preventing alveolar collapse, increasing functional residual capacity, and reducing the work of breathing. In neonatal care, it is mainly indicated for initial and post-extubation respiratory support and recurrent apnea in PT infants, and for congenital heart defects with increased pulmonary blood flow, chronic conditions such as bronchopulmonary dysplasia (BPD), neuromuscular diseases, etc. In PT infants, it reduces the use of MV and the incidence of BPD.5,6
Ventilators and interfacesTwo types of devices are used to deliver neonatal NIV. Both should be used with a heated humidifier for adequate gas conditioning:
- •
Continuous-flow devices: they generate positive pressure through expiratory resistance, either using an exhalation valve (conventional ventilators) or by immersion of the distal end of the expiratory limb in water (bubble continuous positive airway pressure [CPAP]).7
- •
Variable-flow devices: specifically designed for NIV. Positive pressure is generated by altering gas flow across the phases of the respiratory cycle. This produces a stable pressure level that reduces the work of breathing.7
The interface is crucial for the success of NIV. Its size must be selected correctly, and it should be fitted tight enough to reduce leaks and loose enough to prevent injury to the nose or face. The evidence supports the use of nasal masks and short binasal prongs (alternating them), although there are other interfaces, such as oronasal masks (used in resuscitation), high-flow nasal cannulas, and cannulas with long and narrow tubing (RAM cannula).8
Modalities, synchronization, and parametersModalities with a single pressure level are known as nCPAP. When it comes to modalities with two pressure levels, there are two types: bilevel CPAP (BLPAP, nBiPAP, SiPAP, biphasic CPAP or duoPAP) and intermittent-pressure NIV, which may be either not synchronized (nasal intermittent positive pressure ventilation [nIPPV]) or synchronized (nasal synchronized intermittent positive pressure ventilation [nSIPPV]). Other modalities include nasal high-frequency oscillatory ventilation (NHFOV) and non-invasive neurally adjusted ventilation assist (NIV-NAVA).9 High-flow nasal cannula (HFNC) uses comfortable interfaces, although the delivered pressure is uneven and imprecise, so it is reserved for mild respiratory failure in PT infants born after 28 weeks of gestation or for weaning off NIV.10
Table 1 presents the recommended settings for each modality.
Initial settings recommended for the different noninvasive ventilation modalities in neonates.
| Modality | Flow | PEEP | PIP | RR | TI | Synchronization | Discontinuation |
|---|---|---|---|---|---|---|---|
| CPAP | Continuous (CF-CPAP): 8−10 L/min Variable (VF-CPAP): Not adjustable | 5−6 cm H20 (initial) 8−9 cm H20 (maximum) | – | – | – | – | ↓ CPAP to 4−5 cm H2O (FiO2 < 0.3) |
| nIPPV | Continuous: 8−15 L/min | 5−6 cm H2O (initial) | 10−15 cm H2O (initial) 20 cm H2O (maximum) | 20−25 breaths/min (initial) 40−45 breaths/min (maximum) | 0.4−0.6 s | Possible (capsule, sensor) | ↓ PIP and RR Switch to CPAP |
| Bilevel CPAP | Variable | 5−6 cm H2O (initial) | Pduo 8–10 cm H2O (ΔP 2−4 cm H20) | 10−15 breaths/min (initial) 20 breaths/min (maximum) | 0.6−1 s | Possible (sensor) | ↓ ΔP Switch to CPAP |
| nHFOV | Continuous: 10−15 L/min | MAP: 10−20 cm H2O (initial) + 2 above PEEP/PIP | Amplitude 30−55 cm H2O | Frequency: 8−12 Hz | 33−50% | – | ↓MAP to 8 cm H20 and switch to CPAP/ NIPPV |
| NIV-NAVA | Variable | 5−6 cm H20 (initial) | NAVA level 1−3 cm H20/μV Edi trigger 0.5 μV Apnea time 2−5 s | 40−60 breaths/min (backup RR) | – | NAVA sensor | ↓ NAVA level |
| HFNC | 4−8 L/min (2 L/kg/m) | – | – | – | – | – | Decrease to 2 L/min |
Abbreviations: breaths/min, breaths per minute; CPAP, continuous positive airway pressure (CF-CPAP, continuous-flow CPAP; VF-CPAP, variable-flow CPAP); Edi, electrical activity of the diaphragm; FiO2, fraction of inspired oxygen; HFNC, high-flow nasal cannula; L/min, liters per minute; MAP, mean airway pressure; NAVA, neurally adjusted ventilatory assist; nHFOV, nasal high-frequency oscillatory ventilation; nIPPV, nasal intermittent positive pressure ventilation; NIV, noninvasive ventilation; PEEP, positive end-expiratory pressure; PIP, peak inspiratory pressure; RR, respiratory rate; TI, inspiratory time; ΔP, pressure difference.
Synchronization is the main challenge in neonatal NIV. It can be achieved by means of a pneumatic abdominal capsule (not exact), flow or pressure sensors, or the electrical activity of the diaphragm (NIV-NAVA), which is not affected by air leaks. The latter modality and nSIPPV seem most effective for respiratory support in infants with apnea of prematurity.11
Noninvasive ventilation failureBefore considering that NIV has failed, it is important to ensure optimal implementation, verifying the correct positioning and patency of the airway, proper functioning of the device, and absence of leaks and condensation in the circuit, in addition to ruling out complications such as air leak syndrome and, in PT infants, assessing the need of surfactant.12
When clinical and/or blood gas objectives are not achieved with NIV, possible options include increasing the median airway pressure or switch to a bilevel modality.13 Nasal high-frequency oscillatory ventilation and NIV-NAVA could be used for rescue in case of repeated failure.11,14 The criteria for defining NIV failure are: failure to achieve the target oxygen saturation or achieving it at the cost of a high (> 0.4) fraction of inspired oxygen (FiO2), a PaCO2 greater than 60 mm Hg with a pH of less than 7.25 (in acute conditions), or more than three apnea episodes requiring vigorous stimulation within a one-hour window.15
Conventional invasive mechanical ventilationIntroductionUp to 20% of admitted newborns require conventional invasive MV during the hospital stay, and this percentage can increase to up to 80% among the most immature infants depending on their gestational age and the underlying condition.16
The main indications include NIV failure as well as specific clinical conditions such as acute respiratory distress syndrome, sepsis/shock, persistent pulmonary hypertension, or neuromuscular disorders that compromise spontaneous breathing.
Lung-protective ventilationThe lungs of newborns, especially PT newborns, are extremely vulnerable due to their structural immaturity, limited functional residual capacity, tendency toward alveolar collapse (associated with a soft, highly compliant chest wall), and marked regional heterogeneity.1
It is now widely accepted that the main mechanisms involved in the development of VILI are not directly related to the delivered pressure, as previously thought, but are primarily associated with changes in lung volume. In this regard, alveolar overdistention secondary to high tidal volumes (volutrauma) and repeated collapse and expansion of unstable alveolar units (atelectrauma) are the most relevant pathophysiological factors. Both mechanisms promote a secondary pulmonary inflammatory response (biotrauma), increasing alveolar-capillary permeability, surfactant inactivation, and structural damage to the immature lung.1,17 In consequence, the goal of intervention, whenever possible, is to keep the time of conventional invasive mechanical ventilation as short as possible using a protective strategy from the outset, with a plan for transitioning to NIV to minimize the risk of developing chronic lung disease.1,17,18
To reduce this risk, protective ventilation strategies should be aimed at minimizing alveolar overdistention and maintaining adequate alveolar recruitment. Recommended measures include the use of synchronized modes with low and stable tidal volumes (VT), adjustment of the positive end-expiratory pressure (PEEP) to a level sufficient to reduce cyclic collapse, and programming appropriate inspiratory and expiratory times to minimize air trapping.17,18
SynchronizationSynchronization is a key factor in optimizing patient-ventilator interaction during neonatal conventional invasive MV. In synchronized ventilation, the ventilator detects the spontaneous inspiratory effort of the infant and delivers support by adapting to the infant’s respiratory cycle, rather than delivering ventilations independently at fixed intervals. Effort can be detected through flow or pressure signals generated in the airway, or though neural signals in NIV-NAVA. This approach improves patient-ventilator synchrony, reduces the work of breathing, and contributes to lowering the risk of lung injury associated with asynchrony, promoting more stable and physiological ventilation.19,20
Volume guaranteeThe most relevant advance in neonatal conventional invasive MV in recent years has been the introduction of a ventilation mode that allows setting a target VT or volume guarantee (VG).18 These strategies involve automated control algorithms that adjust the inspiratory pressure dynamically, breath to breath, in order to deliver the precise amount of gas required to achieve the preset VT. In the newly proposed ventilation taxonomy, this is referred to as an “adaptive targeting scheme”.19 This is particularly useful in newborns, in whom lung compliance, airway resistance and inspiratory effort can change substantially during respiratory support, so that the same pressure can give rise to very different tidal volumes. The evidence to date shows that the use of VG reduces exposure to excessive or insufficient tidal volumes and attenuates the pulmonary inflammatory response, and it is associated with fewer days of conventional invasive MV, a lower incidence of pneumothorax and intraventricular hemorrhage grade 3 or 4, a reduction in hypocapnia and a reduction in the composite outcome of death or BPD.21,22
From a practical standpoint, the target VT is set between 4 and 6 mL/kg to optimize gas exchange while minimizing the risk of volutrauma. It should be noted that a reduction in the work of breathing has been reported with targets of at least 5 mL/kg in PT infants with acute respiratory distress syndrome (ARDS), whereas patients with evolving BPD may require higher volumes (7−8 mL/kg) due to a greater physiologic dead space.23 These targets should be interpreted in the context of the exhaled VT, leaks, ventilator curves, and the patient’s clinical condition, as the precision of the VT depends on the ventilator, the measurement location and the magnitude of the leaks.
Modalities and clinical conditionsAlthough volume-controlled devices capable of delivering very small volumes (therefore, appropriate for use in newborn infants) have been developed in recent years, most neonatal intensive care units (NICUs) have pressure-controlled devices. This modality is characterized by a decelerating inspiratory flow, a limited inspiratory pressure and, usually, ventilation cycles set according to the inspiratory time.
Within pressure-control, there are different available ventilation modalities that differ mainly in the targeting scheme and the type of breath assist-control (AC; PC-CMVs), synchronized intermittent mandatory ventilation (SIMV; PC-IMV (1)s), synchronized intermittent mandatory ventilation with pressure support (SIMV with PS; PC-IMV (1)1 s,s), other variants of IMV such as Pressure Support Ventilation (PSV) on Dräger machines (PC-IMV(2)s,s) or continuous spontaneous ventilation/pressure support (PSV; PC-CSVs). Used alone, SIMV (PC-IMV(1)s) delivers a fixed number of synchronized mandatory ventilations and does not deliver pressure support for the intervening spontaneous breaths which, in neonatal patients, increases the work of breathing and may prolong weaning, so this mode has been largely replaced by techniques that support all breaths or combine SIMV with PS (PC-IMV (1)1 s,s) and is not generally recommended as a first-line ventilation strategy.24,25 Recent evidence suggests that in SIMV (VG) with PS (PC-IMV(1)a,s), the lack of automatic adjustment of the PS level for spontaneous breaths to the automatically controlled pressure delivered in mandatory inflations makes it difficult to achieve stable tidal volumes, so this strategy would be less protective compared to other modalities.25
Overall, neonatal conventional invasive MV should be delivered with synchronized and, preferably, VG modalities to ensure lung protection.18,24 Settings should be individualized according to the particular respiratory condition and stage of disease of the patient. For general guidance, we suggest initial settings for the most common conditions encountered in neonatal care (Table 2).
Initial settings recommended for the most common conditions in the neonatal period.
| Clinical condition | Main goal | Initial settings | Key points |
|---|---|---|---|
| RDS (surfactant deficiency) | Protective ventilation Prevention of atelectasis | PEEP 5−8 cm H2O TI 0.3–0.35 s High RR: 50−55 breaths/min | Reevaluate after surfactant administration. Avoid overdistention and fluctuations in PaCO2. Short CT |
| Preterm birth (GA < 28 wk) | Minimizing risk of VILI | PEEP 5−6 cm H2O TI 0.30-0.35 s RR 40−60 breaths/min | Ventilation with VG Permissive hypercapnia Appropriate use of PEEP |
| Pneumonia Sepsis | Optimizing gas exchange | PEEP 5−7 cm H2O TI 0,30–0,40 s RR 20−40 breaths/min | Risk of auto-PEEP with high RRs; synchronization to reduce work of breathing |
| Perinatal asphyxia HIE | Stable ventilation | PEEP 5−6 cm H2O TI 0,35-0,40 s RR 20−40 breaths/min | Minimize PaCO2 fluctuations Avoid hyperventilation |
| Air leak syndrome | Minimizing leak Support | Minimum necessary PIP/PEEP Short TI High RR: 40−55 breaths/min VT < 6 mL/kg | Low compliance. Short CT |
| Meconium aspiration syndrome | Optimizing oxygenation Preventing air trapping | Restrictive pattern PEEP: 5−10 cm H2O VT: 5.5-6 mL/kg Short CT: RR, 50 breaths/min Obstructive pattern PEEP: 4−5 cm H2O VT: 5 mL/kg Long CT: RR, 30 breaths/min; prolonged TE (0.7-1 s) | Watch for hyperinflation and auto-PEEP Monitor flow curves |
| HTPP | Optimizing oxygenation and reducing PVR | PEEP 5−6 cm H2O FiO2 for PaO2: 70-100 mm Hg | Avoid overdistention (makes hemodynamic parameters worse) maintain stable PaCO2 |
| Established BPD | Ensuring effective ventilation in a heterogeneous lung | PEEP 6−8 cm H2O PIP required for VT VT 6-8 mL/kg TI 0.35-0.45 s RR 20−40 breaths/min | Watch for auto-PEEP. Prolonged expiratory times Long CT |
Abbreviations: BPD, bronchopulmonary dysplasia; breaths/min, breaths per minute; FiO2, fraction of inspired oxygen; GA, gestational age; HIE, hypoxic-ischemic encephalopathy; PEEP, positive end-expiratory pressure; PPHN, persistent pulmonary hypertension of the newborn; PVR, pulmonary vascular resistance; RDS, respiratory distress syndrome; RR, respiratory rate; TC, time constant (τ); TE, expiratory time; TI, inspiratory time; VG, volume guarantee; VILI, ventilation-induced lung injury; VT, tidal volume.
High-frequency oscillatory ventilation emerged in the late 1970s as a promising lung-protective strategy based on the use of tidal volumes smaller than the anatomical dead space at supraphysiologic frequencies (expressed in hertz [Hz], with 1 Hz = 60 breaths/min) and maintaining a constant distension pressure on the airway (based on mean airway pressure [MAP]) in order to minimize VILI.
Today, it is chiefly used for rescue in patients with respiratory failure refractory to conventional invasive MV. However, the most recent systematic reviews26–28 have found a small but significant reduction in the incidence of BPD with elective HFOV compared to conventional MV in trials that implemented a strict lung volume recruitment strategy27 or included administration of surfactant28 in preterm infants with RDS. Other conditions in which HFOV can be beneficial include air leak syndromes (pulmonary interstitial emphysema or pneumothorax), persistent pulmonary hypertension of the newborn refractory to conventional invasive MV, pulmonary hemorrhage, and atelectasis.
Settings and managementIn HFOV, ventilation depends on the high-frequency tidal volume (VThf), which results from the set amplitude and frequency, in addition to the condition of the lungs. Both parameters are comprehended in the carbon dioxide diffusion coefficient (DCO2) according to the following formula: DCO2 = VThf2 × HFV frequency.
In recent years, thanks to the introduction of VG in HFOV devices, it has become possible to maintain a more stable VThf, reducing fluctuations in ventilation and oxygenation.29,30 Using the highest possible frequencies, close to the resonance frequency, is recommended in order to optimize gas transport and reduce VILI.31 On the other hand, oxygenation depends on lung recruitment, which is determined by the MAP.
Table 3 presents the initial settings for HFOV. The settings are then individualized according to the clinical goals and blood gas targets established for each patient as shown in Table 4.
Initial settings for HFOV.
| Initial settings | ||
|---|---|---|
| Without VG | With VG | |
| FiO2 | Same as invasive MV: based on SpO2 target | |
| MAP | 1−2 cm H2O > MAP in invasive MV; air leak ( = MAP) Lung recruitment maneuver | |
| f | 8−15 Hz Inversely proportional to weight | |
| ΔPhf | No set limit to achieve the target VThf | 20−30 cm H2O, to achieve symmetry of vibrations |
| VThf | Resulting from set ΔPhf and f | 1.5−2 mL/kg |
| TI | I:E 1:2 (33%) | |
Abbreviations: f, frequency; FiO2, fraction of inspired oxygen; I:E, inspiratory-to-expiratory ratio; MAP, mean airway pressure; MV, mechanical ventilation; TI, inspiratory time; VG, volume guarantee; VThf, high-frequency tidal volume; ΔPhf, change in pressure.
Maneuvers to modify oxygenation and ventilation in HFOV, with and without volume guarantee.
| Oxygenation | Ventilation | ||
|---|---|---|---|
| Lung recruitment | DCO2 = VThf2 × f | ||
| MAP FiO2 | Without VG | With VG | |
| ↑ΔPhf ↓ f | ↑VThf | ↑VThf ↑f | |
Abbreviations: DCO2, carbon dioxide diffusion coefficient; f, frequency; FiO2, fraction of inspired oxygen; MAP, mean airway pressure; VG, volume guarantee; VThf, high-frequency tidal volume; ΔPhf, change in pressure.
Acute respiratory distress in preterm infants
It is a form of acute respiratory failure caused by a quantitative and qualitative deficiency of surfactant in the immature lung. The lack of surfactant increases alveolar surface tension, which promotes alveolar collapse, leading to atelectasis and intrapulmonary shunting. This decrease in lung compliance results in a reduced functional residual capacity, which in turn causes progressive respiratory distress with hypoxemia, accompanied, to a lesser extent, by hypercapnia.
Its incidence is inversely correlated to gestational age, and other factors, such as maternal diabetes, prolonged rupture of membranes or perinatal asphyxia, are also associated with an increased risk. It is diagnosed based on the clinical presentation supported by radiographic and/or sonographic findings.2,19
The management must be approached as a stepwise selection of lung-protective strategies aimed at avoiding intubation whenever possible, with CPAP/NIV as the first-line modes of respiratory support in addition to early administration of surfactant in moderate to severe cases, preferably by minimally invasive techniques (less invasive surfactant administration [LISA]/ minimally invasive surfactant therapy [MIST]) to reduce exposure to conventional invasive MV and the risk of complications such as pneumothorax and, according to some studies, intraventricular hemorrhage32,33 (Fig. 1). Intubation and conventional invasive MV are reserved for cases of NIV failure despite optimization, always with implementation of lung-protective strategies and an early weaning protocol19 (Fig. 2).
Respiratory instability in preterm infant.19
Abbreviations: CPAP, continuous positive airway pressure; FiO2, fraction of inspired oxygen; Fr, respiratory frequency; LISA, less invasive surfactant administration; LUS, Lung Ultrasound Score; NICU, neonatal intensive care unit; nIPPV, nasal intermittent positive pressure ventilation; PaCO2, partial pressure of carbon dioxide in arterial blood; PaO2, partial pressure of oxygen in arterial blood; PEEP, positive end-expiratory pressure.
Invasive respiratory support for management of neonatal respiratory distress syndrome in preterm infants after NIV failure.
Abbreviations: FiO2, fraction of inspired oxygen; HFOV, high-frequency oscillatory ventilation; NIV, noninvasive ventilation; nIPPV, nasal intermittent positive pressure ventilation; MV, mechanical ventilation; pCO2, partial pressure of carbon dioxide; PIP, peak inspiratory pressure.
The criteria for definition of NIV failure are defined above.
This condition is defined as the presence of air outside the alveoli and is one of the main complications of conventional invasive MV, with an incidence of 6% to 10% in ventilated PT infants. Air leaks are classified according to the location of the air as pneumothorax, pneumomediastinum, or interstitial emphysema. Air leaks are associated with alveolar overdistention in conditions presenting with lung heterogeneity, which eventually results in tissue rupturing and passage of air to the interstitial and/or pleural space. Clinically, it should be suspected in the event of a sudden deterioration in oxygenation or increase in work of breathing with asymmetric air entry, which can progress to hemodynamic instability in tension pneumothorax. It can be diagnosed rapidly by point-of-care ultrasound or plain radiography.34
The treatment combines a lung protective strategy with limitation of pressure and volume (reducing the PIP/MAP and optimizing times [short time constant], allowing permissive hypercapnia and titrating sedation and analgesia), and HFOV can be considered in the case of interstitial emphysema or persistent air leak.19
Specific interventions are guided by the severity of the condition: individualized management (observation or needle aspiration) for simple pneumothorax, based on the symptoms, tube drainage in the case of tension pneumothorax or persistent air leak, and rescue maneuvers such as selective intubation of the healthy lung in refractory cases.
Noninvasive ventilation is relatively contraindicated in patients with an undrained pneumothorax
Meconium aspiration syndromeIt is a form of neonatal respiratory distress caused by perinatal aspiration of meconium-stained fluid, leading to chemical pneumonitis with inactivation of surfactant and airway obstruction (usually with a ball-valve effect with air trapping and an increased risk of air leak syndromes).35 Due to the uneven distribution of meconium, the same patient can exhibit a restrictive pattern if atelectasis develops secondary to total bronchial obstruction and an obstructive pattern with hyperinflated areas due to air trapping. These processes collectively worsen ventilation-perfusion mismatch, leading to hypoxemia, CO2 retention, and development of mixed acidosis.19,35
It should be suspected in neonates (usually full-term) with a history of meconium-stained fluid at birth presenting with respiratory distress and progressive hypoxemia and with compatible imaging findings.35
Respiratory support must be individualized according to the pattern that predominates at the moment (obstructive or restrictive), blood gas targets, and the observed lung mechanics. The current evidence does not support the superiority of a specific conventional invasive MV strategy compared to the others.19 Given the patchy involvement of the lungs in meconium aspiration syndrome, this condition can benefit from volume-control ventilation, as delivery of a constant flow and the distribution of volume during the inspiratory pause can be useful in lung diseases with heterogeneous involvement.24,25,35 High-frequency oscillatory ventilation is an alternative option when adequate oxygenation cannot be achieved in cases with a markedly restrictive pattern.31
Adjunctive therapies can also be used:
- none-
Sedation: sedation is beneficial in most severe cases, and muscle blockers may be useful in some cases.35,36
- none-
Surfactant: its use remains controversial,37–41 although it has been growing, especially in patients with moderate to severe disease, and particularly in cases of CPAP failure requiring conventional invasive MV or with persistent hypoxemia. It can be administered as a bolus,39 as surfactant lung lavage,40 or with less invasive methods, such as LISA/MIST.42
- none-
In the case of concomitant pulmonary hypertension, maintaining a PaO2 of 60 to 80 mm Hg is recommended, also considering the use of inhaled nitric oxide (iNO) and/or sildenafil.43
It is a chronic manifestation of lung injury in preterm infants that results from the anatomical immaturity of the lungs and the environment in which they mature (exposure to oxygen, respiratory support, nutritional status, intercurrent infection).2 Its incidence is as high as 40% in PT infants born before 28 weeks and has barely decreased in the past few decades.
It is characterized by persistent inflammation and disruption of alveolar and vascular development resulting in alveolar simplification, fibrosis, an increase in dead space and a reduction in the surface area available for gas exchange. Its functional impact is variable, as it can give rise to restrictive patterns due to collapse or obstructive patterns due to hyperinflation/air trapping. Its association with pulmonary hypertension is variable.19
Preventive intervention is key, with optimization of the baseline condition of the lungs (antenatal steroids) and implementation of lung-protective strategies from birth: support with NIV, early surfactant administration (LISA), avoiding hyperoxia, and use of VG.44
Once BPD develops, given its clinical heterogeneity, respiratory support must be individualized, keeping in mind that greater tidal volumes (6−8 mL/kg) may be necessary due to the increase in dead space.44
With regard to pharmacological treatment, only low-dose dexamethasone administered beyond the 7th day of life has been associated with earlier extubation, without significant short-term side effects. There is consensus regarding the importance of optimizing nutritional status and preventing infection. Other adjunctive therapies, such as diuretics, methylxanthines, vitamin A, bronchodilators, and pulmonary vasodilators, play a limited role in symptom management.
ConclusionsThe immaturity of newborn infants, especially those born preterm, results in physiological particularities that make them highly dependent on respiratory support and, at the same time, particularly vulnerable to its adverse effects. Understanding these characteristics is essential for optimizing the use of ventilation and minimizing the risk of lung and brain injury. In this context, the use of individualized, lung-protective, and synchronize ventilation strategies must be prioritized. Noninvasive ventilation (NIV) is the first-line respiratory support modality, with invasive ventilation reserved for cases of clearly defined NIV failure. Furthermore, synchronization and the use of VG are key factors in reducing ventilation-induced lung injury and improving both short- and long-term clinical outcomes.
Declaration of Generative AI and AI-assisted technologies in the writing processDuring the preparation of this work the authors used ChatGPT (OpenAI) and Copilot in order to edit previously written paragraphs for the sole purpose of improving readability. After using these tools, the authors reviewed and edited the content as needed and take full responsibility for the content of the publication.